Water TS International Co., Ltd
+8615809208763

Which instruments are usually required for reverse osmosis systems?

Apr 06, 2022

1) SDI15 instrument: SDI15 value is the sludge deposition index, which is a very important detection index. The reverse osmosis influent requires SDI15<5.

Consider using an economical manual SDI meter with periodic manual inspection. Large projects can also use fully automatic SDI instruments without bad money.

2) Oxidation-reduction potential meter (ORP): The reverse osmosis membrane is not resistant to chlorine oxidation, and the influent water needs to remove residual chlorine, which can be adsorbed by adding a reducing agent or activated carbon. The removal effect can be detected by the ORP table, and the reverse osmosis influent requires that the ORP is usually controlled within +200mV.

RO system

3) pH meter: In order to protect the operation performance of RO membrane, for the pH adjustment process or industrial wastewater water quality in the previous stage, a pH meter is added to detect the pH value of reverse osmosis influent, and upper and lower alarm values are set.  

4) Thermometer: The temperature has a great influence on the water production of the reverse osmosis membrane, and a thermometer is set to detect the temperature change of the incoming water online to facilitate standardization.  

5) Pressure gauge: pressure gauges are installed at the inlet and outlet of the security filter to judge the operation status of the filter element;  

The outlet of the high-pressure pump is provided with a pressure gauge and a high-pressure switch, and the inlet is provided with a low-pressure switch; 

 Each section of the reverse osmosis unit is equipped with a pressure gauge to record the pressure drop of each section to determine whether the membrane element is polluted or fouled;  

6) Pressure transmitter: each section of the reverse osmosis device is equipped with a pressure transmitter to record the pressure drop of each section online; 

RO system instruments 2

7) Flow meter: The reverse osmosis device needs to be equipped with a flow meter of produced water and concentrated water to adjust the recovery rate and check whether the produced water is normal.

8) Conductivity meter: A conductivity meter is set up for the influent and produced water to judge the desalination rate of the RO unit.